  2. 国外雑誌論文

In Vivo Femoral Head Damage and Its Effect on Polyethylene Wear

https://asahikawa-med.repo.nii.ac.jp/records/2055
https://asahikawa-med.repo.nii.ac.jp/records/2055
6ce2b724-f1fe-4375-a16b-2e12464427b5
名前 / ファイル ライセンス アクション
2820.pdf 2820.pdf (438.6 kB)
Item type 学術雑誌論文 / Journal Article_02(1)
公開日 2010-02-01
タイトル
タイトル In Vivo Femoral Head Damage and Its Effect on Polyethylene Wear
言語 en
言語
言語 eng
資源タイプ
資源タイプ journal article
著者 伊藤, 浩

× 伊藤, 浩

ja 伊藤, 浩

ja-Kana イトウ, ヒロシ

Search repository
Maliney, CM

× Maliney, CM

en Maliney, CM

Search repository
Crowninshield, RD

× Crowninshield, RD

Crowninshield, RD

Search repository
Clohisy, JC

× Clohisy, JC

en Clohisy, JC

Search repository
McDonald, DJ

× McDonald, DJ

McDonald, DJ

Search repository
Maloney, WJ

× Maloney, WJ

en Maloney, WJ

Search repository
著者 ローマ字
Ito, Hiroshi
著者 ローマ字
en
書誌情報 Journal of Arthroplasty

巻 25, 号 2, p. 302-308, 発行日 2010-02-01
ISSN
収録物識別子タイプ ISSN
収録物識別子 0883-5403
DOI
識別子タイプ DOI
関連識別子 10.1016/j.arth.2009.01.010
リンクURL
内容記述タイプ Other
内容記述 http://www.ncbi.nlm.nih.gov/pubmed/19201153 | http://www.ncbi.nlm.nih.gov/pubmed/19201153
抄録
内容記述タイプ Abstract
内容記述 The purposes of this study were to determine the spectrum of femoral head damage in patients undergoing revision total hip arthroplasty and to determine the impact of that damage on polyethylene wear. One hundred eight consecutive modular metal femoral heads were retrieved at revision surgery. The mean roughness (Ra) value was 0.18 ± 0.18 μm. The roughest femoral heads (mean Ra, 0.56 μm) were from retrievals correlated with mode 2 wear (recurrent dislocation and complete wear through of the polyethylene liner). Five million cycles of wear tests were performed using retrieved femoral heads against both new conventional and highly cross-linked polyethylene. The mean wear rate of conventional polyethylene was 15.9 ± 4.3 mg and that of highly cross-linked polyethylene was 0.04 ± 0.14 mg per 1 million cycles (P < .001). Highly cross-linked polyethylene was more resistant to wear than conventional polyethylene, even when mated against roughened femoral heads.
言語 en
資源タイプ
内容記述タイプ Other
資源タイプ text
フォーマット
内容記述タイプ Other
内容記述 application/pdf
ID(XooNIps)
19201153
閲覧数(XooNIps)
ダウンロード数(XooNIps)
865
戻る
0
views
See details
Views